Home > Incubators > CO2 Incubators > Model 51026282
$9,278.00
» Standard Lead Time: 10-15 Days
Jump To...
Specifications
Condition |
New, Factory-Direct |
Brand |
Thermo Scientific |
Part No. |
51026282 |
Temp Range |
Ambient + 3 to 55°C |
Volume |
150 Liters |
Uniformity |
±0.3°C at 37°C |
Accuracy |
±0.1°C at 37°C |
Chamber |
23.9"H x 18.5"W x 20.9"D |
Shipping Weight |
195 lbs |
Dimensions |
34"H x 25"W x 30.2"D |
Accessories |
3 Shelves |
Electrical |
120 V / 60 Hz |
Product Details
Thermo Scientific Heracell 150i CO2 incubators come standard with an on-demand, automatic 90°C decontamination function that decontaminates the entire chamber interior without the need to remove sensors, fittings and shelves. Decontamination is highly effective against a wide range of contaminants, including mycoplasma.
Electropolished stainless steel interior
Thermo Scientific iCAN touchscreen places total control and complete information at your fingertips
Unique pan-less humidity water reservoir provides faster humidity recovery, promoting improved cell growth
Shelves are perforated, stainless steel
Humidity: RH Ambient to 95% @ 37°C
Includes Access Port with Filter 1.1/4 in. (32 mm) with removable silicone plug
CO2 Inlet: 1/8 in. hose (barbed)
iCAN Touchscreen improves your visibility and control of important incubator information helping you to achieve your culturing goals
Simplifies operation and enables rapid access of important information for each critical parameter in the incubator
Provides trend analysis for convenient evaluation of performance
Screen is door mounted for easy accessibility and viewing
Select from a variety of languages
Visibility to changes in culture environment: on screen logs and usage recording view changes in culture environment etc
Monitor alarm alerts visually on the display
On board data logging with optional USB output
Patented Humidity System for Faster Recovery Rates
Incorporates a unique integral humidity water reservoir that provides a high relative humidity (rH) and allows rapid recovery of optimal humidity level after door openings
A water level sensor indicates when a refill is needed - via a convenient prompt on the touchscreen display - to avoid the desiccation of important cultures
Pan-less system reduces handling and provides recovery rates up to five times faster than ordinary tray humidified incubators
Surface area larger than ordinary humidity water pans (provided by a water reservoir with inclined and rounded corners)
Patented floor heating system that operates after door opening
Direct heat-transfer from heated floor to humidity reservoir
CO2 Range of 0-20%
CO2 Control accuracy: Better than ±0.1%
CO2 Sensor: Long Life Thermal Conductivity (TC) Sensor
CO2 Readability and Setability: 0.10%
CO2 Inlet pressure required: 15 PSIG (1.0 bar)
Included Accessories
3 Shelves
Available Accessories